Some ingredients contained in the medicines are liable to be oxidized, which may cause the deterioration of the medicines. That’s why the packaging materials used for medicines are required to have sufficient oxygen barrier properties. Considering the drug safety, the packaging materials for medicines should meet specific requirements especially for the oxygen barrier property. The packaging for medicines shall have sufficient oxygen barrier property so that the medicines can be protected from oxidization effectively. The common packaging materials for medicines include plastic composite films, aluminized plastic films and aluminum plastic composite films, etc. Among all those materials, aluminum plastic composite film is featured with high barrier propertymeanwhile the packaging using such material usually has more pleasant appearance. However, the aluminum foil layer is liable to be folded during the process of packaging molding, drug filling, heat sealing, transporting and selling. Once the aluminum foil layer is folded, the barrier property of the composite film may be compromised and the seal performance of the packaging will be decreased adversely.
The oxygen barrier property of the packaging is one main factor that may influence the shelf life of the medicines. To be specific, if the package for medicines is of poor oxygen barrier property, the medicines may be oxidized, deteriorated and lose efficacy especially after long-term transportation and storage. Therefore, it is necessary for the drug manufacturers to strengthen the regular monitoring of the oxygen barrier property of drug packaging.
This article provides a detailed testing solution of oxygen barrier property if aluminum plastic composite packaging for powder-form medicines, which may be helpful to those medical companies that have testing requirements for oxygen barrier properties. The testing instrument we used is Labthink’s VAC-V2 Gas Permeability Tester.
The tests can be performed according to the following procedures.
Cut 3 pieces of specimens with sample cutter. The diameter of the specimen should be F97 mm.
Mount the 3 specimens onto 3 test cells of the testing instrument.
Set specimen name, specimen thickness, test temperature, test humidity and other parameters.
Turn on the vacuum pump. Click Start Test to start the test. The test results will be calculated and displayed automatically.
For more details about VAC-V2 Gas Permeability Tester, please visit www.labthink.com
About Labthink Instruments Co., Ltd:
Labthink Instruments Co., Ltd is one leading supplier of packaging testing instruments, which is headquartered in Jinan, China and operated in Boston, U.S.A.